pHotoFlex / pHotoFlex Turb
Operation
57
ba75491e05
04/2009
Sample printout of a
record
Calibration evaluation
After calibrating, the meter automatically evaluates the calibration. The
asymmetry and slope are evaluated separately. The worse evaluation
of both is taken into account. The evaluation appears on the display
and in the calibration record.
31.10.03 16:13
pHotoFlex
Ser. no.
12345678
Calibration
pH & ORP
Calibration date
31.10.03 16:13:33
Calibration interval
7 d
AutoCal TEC
Buffer
1 4.01
Buffer
2 7.00
Buffer
3 10.01
Voltage
1 184.0 mV 24.0 °C
Voltage
2 3.0 mV 24.0 °C
Voltage
3 -177.0 mV 24.0 °C
Slope
-60.2 mV/pH
Asymmetry 4.0 mV
Sensor
+++
Sample dis-
play
Calibration
record
Asymmetry
[mV]
Slope
[mV/pH]
+++
-15 ... +15
-60.5 ... -58
++
-20 ... +20
-58 ... -57
+
-25 ... +25
-61 ... -60.5
or
-57 ... -56
-
-30 ... +30
-62 ... -61
or
-56 ... -50
Clean the electrode according
to the electrode operating
manual
----
----
< -30 or
> 30
... -62 or
... -50
Eliminate the error according to
chapter 6 W
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...