![wtw pHotoFlex Operating Manual Download Page 74](http://html1.mh-extra.com/html/wtw/photoflex/photoflex_operating-manual_1000666074.webp)
Operation
pHotoFlex / pHotoFlex Turb
74
ba75491e05
04/2009
Socket assignment
4.9.2
Configuring the RS232 interface
For an error-free data transmission, the RS232 interface should be set
to the same transmission speed (Baud rate) on the pHotoFlex (Turb)
and PC/printer.
You can set the following values for the baud rate on the
pHotoFlex (Turb) : 1200, 2400, 4800, 9600, 19200.
The baud rate is selected in the menu, Configuration / System / Inter-
face / Baud rate.
4.9.3
Selecting the output format of datasets
For downloading data to the interface you can select the output format.
It is selected in the menu, Configuration / System / Interface / Output
format.
The ASCII output format delivers formatted datasets.
The CSV output format delivers datasets separated by ";".
Output format, ASCII
RS 232
REC
1
3
4
5
6
2
1 RTS
2 RxD
3 TxD
4 DTR
5 SG
6 CTS
pHotoFlex
Ser. no.
12345678
31.10.04 09:56:20
ID: 1
10.01 (pH) AR
25 °C
________________________________________
pHotoFlex
Ser. no.
12345678
31.10.04 15:48:08
ID 1 / PROG 2
1.1 mg/l Pb [BV][9 + 1]
________________________________________
etc...
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...