Troubleshooting
Troubleshooting
226
BA61101/08.00/Pla/MLP5-E2
BA61101/08.00/Pla/MLP5-E2
227
BSB5
Cal Error
BSB5 Anf.mess.
Initial measurement:
Missing calibration data.
- Possible cause / Action:
- RESET performed / Calibrate
- Initial setup / Calibrate
Cal Error
BSB5 Endmess.
Final measurement:
Missing calibration data.
- Possible cause / Action:
- RESET performed / Calibrate
- Initial setup / Calibrate
Ãmeth.
BSB5
BOD5 aktiv
Procedure BSB5 measurement:
Databank set up for measurements
according to the ISO standard.
Action: Erase databank.
ATTENTION, all data of BOD5
measurements will be lost!
BOD
Cal Error
BOD5 time zero
Initial measurement:
Missing calibration data.
- Possible cause / Action:
- RESET performed / Calibrate
- Initial setup / Calibrate
Cal Error
BOD5 after 5d.
Final measurement:
Missing calibration data.
- Possible cause / Action:
- RESET performed / Calibrate
- Initial setup / Calibrate
Ãmeth.
BOD5
BSB5 active
Procedure BOD5 measurement:
Databank set up for measurements
according to the DIN standard.
Action: Erase databank.
ATTENTION, all data of BSB5
measurements will be lost!
Consumption rate
Cal Error
cons.
Measurement: Missing calibration data.
Possible cause / Action:
- RESET performed / Calibrate
- Initial setup / Calibrate
Oº < 1m¥/L
Âmeth.
Oº-rate:
Measurement: Initial value below
1 mg/l.
Cause / Action:
- Sample solution not sufficiently aired
/ Air sampe solution and measure once
again
time-out
Âmeth.
Oº-rate:
Measurement: There is no data usable
for evaluation.
Action: Air sample solution and measured
once again.