Documentation
Documentation
174
A61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
175
Oxygen independent on the position of selector switch!
Command Function
CCON
Measure the oxygen concentration
O2 22.02.94 09:19:40 * 8.6 mg/L 21.9 °C
MLP5 >
CPO2
Measurement of oxygen partial pressure
O2 22.02.94 09:19:40 * 168 mbar 21.9 °C
MLP5 >
CSAT
Measurement of saturation
O2 22.02.94 09:19:40 * 107 % 21 °C
MLP5 >
RCAL
Read out the calibration data
OxiCal 01.01.00 00:00:00 1.000 Srel 20.0 °C
MLP5 >
RBNR
Read-out of the BOD5 samples
22.02.94/01
MLP5 >
RBPR
sample
Read-out of BOD5 -analysis for sample. Format see
BOD5- measurement.
Error message
INVALID COMMAND Wrong or not interpretable command
Print format of test values
Printer 80 characters/line or interface command format
3 characters serial number or identification no.
(is dropped in case of interface command)
1 space
(is dropped in case of interface command)
5 characters method indication
1 space
17 characters date and time
1 space
3 characters special signs: ! (for sample blank value or offset
correction), F (for slope correction),* (for measurements out of the
meas. range)
1 space
9 characters test value
1 space
8 characters unit
1 space
5 characters indication form
1 space
4 characters temperature (not for photometric measurements)
1 space
2 characters °C (not for photometric measurements)
Data transmission
Baud rate
1200
Data bits
8
Stop bits
1
Parity
none
Handshake
Hardware
max. cable length
15 m
Font
IBM, separation by blanks