Oxygen measurement
112
BA61101/08.00/IWa/MLP5-E2
Wait for
AutoRead.
T= 21.6°C
ÂOº
S= 0.983
¡AR
Wait until
¡
appears before
the
AR
.
Calibration finished (signal
sound). The determined
data are stored.
Ãmeth.
Âmenu
ÀCal
ÁOº m¥/L 21.6°C
AR
Back to menu oxygen
measurement.
Measurem. of oxygen concentration in water
Ãmeth.
Âmenu
ÀCal
ÁOº m¥/L 21.6°C
AR
Menu oxygen
measurement.
Immerse oxygen probe into sample solution !
Care for incident flow!
Âmenu
Oº m¥/L 21.5°C
AR
7,3
Start measurement.
AR
= Measurement with AR
is running.
Interruption is possible by
changing to menu or
actuation of selector switch.
Âmenu
ÀCal
ÁOº m¥/L 21.4°C
*AR
7,1
Wait until
¡
appears before
the
AR
. Measured values
stored (signal sound).
No
¡
: AutoRead not
reached.
Documentation of meas. values: refer to chapter documentation.