pH measurement
pH measurement
66
BA61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
67
ÂpH
AutoCalDIN ¡
Cd1= 6.865
Cd2= 4.006
Indication of used buffer.
À
and
Á
switches to-and-fro
between the calibration
data.
Ãmeth.
Âmenu
ÀCal
ÁpH
25.4°C
AR TP
Back to pH measurement.
One and two point calibration
ConCal with two freely set buffer pH values
ConCal, the conventional calibration is done manually, i.e. the drift
control AR is automatically switched off.
AutoCalDIN
ÂpH
ConCal
¡AutoCalTEC
AutoCalTEC
ÂpH
AutoCalDIN
¡ConCal
Select calibration procedure
ConCal.
Rinse pH electrode and temperature probe and
immerse into neutral buffer solution.
à TP
ConCal ASY pH
T = 23.6°C
pH=
Î.ÑÏ
Start measurement.
Wait for stable test values.
à TP
ConCal ASY pH
T = 23.6°C
pH=
Ï.ÈÈ
Adjust pH value for the
measured temperature.
ÃSLO
ÂpH
S= -59.2mV/pH
Uasy= -1mV
Theoretical or last with
another calibration mode
determined slope and
measured asymmetry.
Â
: End of the one point calibration and back to measurement.