Conductivity measurement
Documentation
100
BA61101/08.00/Pla/MLP5-E2
BA61101/08.00/Pla/MLP5-E2
153
Set cell constant C and measuring range
Possible measuring ranges:
Measuring range
Recommended measuring cells
LR001/T LR01/T TetraCon
®
96 TetraCon
®
L
0.000 - 3.000
µS/cm
þ
þ
0.00 - 30.00 µS/cm
þ
þ
0.0 - 300.0 µS/cm
þ
þ
þ
0 - 3000 µS/cm
þ
þ
0.00 - 30.00
mS/cm
þ
þ
0.0 - 300.0 mS/cm
þ
þ
0 - 70 (Salinity)
þ
þ
Ãmeth.
Âmenu
Á® 25.4°C
nlF 25
Method conductivity
measurement.
TC nlF
®
documentation
¡C /cm
0.609
Menu "additional functions"
for conductivity
measurement.
C: Select cell constant.
Âmenu
¡C /cm
È.ÎÈÑ
Ãmeas.range
Entry to setting of cell
constant.
Printing of results, function "Ident. no."
switched on
Insert vial.
Menu " Set ident. no." is
displayed.
Set the ident. no.
e.g. no. 184.
The instrument assigns the
set sample code to the last
measured sample.
Display of the test value.
Transmission of data in the
desired adjustments to the
printer.
e.g. printout of 80 characters per line:
(Sequence:Sample number - Method - Date - Time -
Special signs - Test value - Dimension - Indic. form)
001 C1/25 30.04.93 17:46:32 83 mg/L CSB
002 C2/25 03.05.93 09:03:02 1500 mg/L CSB
003 A5/25 04.05.93 09:33:00 !F 1.46 mg/L NH4-N
004 pH 03.05.93 09:07:11 10.75 25.3
°C
005 LF 03.05.93 11:25:23 273.0 mS/cm
006 O2 03.05.93 13:25:25 80.0 mg/L
007 pH 04.05.93 09:25:01 4.83 27.2
°C
008 pH 04.05.93 09:30:00 4.75 27.3
°C
009 pH 04.05.93 09:40:00 5.82 15.2
°C
010 mV 04.05.93 11:11:09 1250 mV 18.2
°C