System 182-XT-4
Installation
3 - 5
ba75753e06
12/2014
Calculation example:
Result: Total power requirement P
<
4.5 W, hence no additional power
supply module is required.
Result: Total power requirement P
>
4.5 W, hence an additional power
supply module is required.
Power supply with the
aid of the
MIQ/Blue PS radio
module
The MIQ/Blue PS radio module can also supply power to components
with a total of up to 7 W power consumption on an IQ S
ENSOR
N
ET
island. Details of this are given in the operating manual of the MIQ/
Blue PS.
3.2.4
Optimum installation of the additional MIQ power supply
module
Basic rule
It is preferable to install the MIQ power supply module in the vicinity of
the IQ S
ENSOR
N
ET
component with the highest power consumption
(example UV/VIS sensor).
Example configuration 1
Power
requirement [W]
(component)
Total power
requirement P [W] (sum
of the components)
+ 1 VisoTurb
®
700 IQ
1.5
1.5
+ 1 ViSolid
®
700 IQ
+ 1.5
3.0
+ 1 FDO
®
700 IQ
+ 0.7
3.7
+ VARiON
® Plus
700 IQ
+ 0.2
3.9
+ 1 DIQ/CHV (ED = 0.1)
(0.8 x 0.1 =) 0.08
3.98
+ 1 DIQ/CHV (ED = 0.05)
(0.8 x 0.05 =) 0.04
4.02
Example configuration 2
Power
requirement [W]
(component)
Total power
requirement P [W] (sum
of the components)
+ 1 FDO
®
700 IQ
0.7
0.7
NitraVis
®
705 IQ
8.0
8.7
MIQ/CHV PLUS (ED = 0.1)
(1.0 x 0.1 =) 0.1
8.8
Summary of Contents for DIQ/S 182-XT-4
Page 6: ...Contents System 182 XT 4 0 4 ba75753e06 12 2014 ...
Page 14: ...Safety System 182 XT 4 2 4 ba75753e06 12 2014 ...
Page 68: ...Installation System 182 XT 4 3 54 ba75753e06 12 2014 ...
Page 86: ...Operation System 182 XT 4 4 18 ba75753e06 12 2014 ...
Page 92: ...The SETTINGS menu System 182 XT 4 5 6 ba75753e06 12 2014 ...
Page 128: ...Maintenance and cleaning System 182 XT 4 7 2 ba75753e06 12 2014 ...
Page 144: ...Technical data System 182 XT 4 9 12 ba75753e06 12 2014 ...
Page 146: ...Accessories and options System 182 XT 4 10 2 ba75753e06 12 2014 ...
Page 150: ...Appendix store separately if required System 182 XT 4 12 2 ba75753e06 12 2014 ...
Page 151: ......