By activating or deactivating the Integral (ti) and Differential (td)
controller part, the following controller types can be configured:
Characteristic curve of
the proportional
controller
For a pure P controller the correlation between the measured value and
current I on the controller output results in the following characteristic
curve:
Fig. 6-13 Proportional controller (negative characteristic curve)
The proportional range Xp is limited by the measurement range of the
linked sensor. If a value is entered for the parameter Xp that is greater
than zero, the controller has a negative characteristic curve (example
Fig. 6-13). To obtain a positive characteristic curve, a negative value
must be entered for Xp.
Controller type
td [s]
ti [s]
P controller
0
0
PI controller
0
1 to 9999
PD controller
1 to 9999
0
PID controller
1 to 9999
1 to 9999
0
Imin
Imax
Current I [mA]
Measured value
Nominal value
Io
Xp
Summary of Contents for DIQ/S 182-XT-4
Page 6: ...Contents System 182 XT 4 0 4 ba75753e06 12 2014 ...
Page 14: ...Safety System 182 XT 4 2 4 ba75753e06 12 2014 ...
Page 68: ...Installation System 182 XT 4 3 54 ba75753e06 12 2014 ...
Page 86: ...Operation System 182 XT 4 4 18 ba75753e06 12 2014 ...
Page 92: ...The SETTINGS menu System 182 XT 4 5 6 ba75753e06 12 2014 ...
Page 128: ...Maintenance and cleaning System 182 XT 4 7 2 ba75753e06 12 2014 ...
Page 144: ...Technical data System 182 XT 4 9 12 ba75753e06 12 2014 ...
Page 146: ...Accessories and options System 182 XT 4 10 2 ba75753e06 12 2014 ...
Page 150: ...Appendix store separately if required System 182 XT 4 12 2 ba75753e06 12 2014 ...
Page 151: ......