Outputs
System 182-XT-4
6 - 20
ba75753e06
12/2014
6.4.7
Sensor controlled
With the Sensor controlled function, the relay is controlled by a linked
sensor.
Requirements
Controller version from 2.80
Sensor that transmits signals to trigger a cleaning cycle, e.g. UV/VIS
sensor
Settings
Setting
Selection/Values
Explanation
Pulse length
Automatic
The duration of the air
cleaning process is
programmed in the
sensor.
The relay automati-
cally takes over the
cleaning duration
from the sensor.
0,5 s
1 s
2 s
3 s
The relay finishes the
cleaning after the
selected interval.
Settings of the cleaning process can be done in the
SETTINGS menu of the respective sensor.