Technical data
System 182-XT-4
9 - 10
ba75753e06
12/2014
Valve circuits
NOTE
The valve may only be operated with the auxiliary voltage of the DIQ/
S 182-XT-4.
Compressed air
Switching voltage
Approx. 22 V
Max. switching current
Approx. 40 mA
Required air quality
Dry, free of dust and oil
Operating pressure
Max. 7x10
5
Pa (7 bar) absolute
Connections on the DIQ/CHV
6 mm hose nozzles