System 182-XT-4
Outputs
6 - 11
ba75753e06
12/2014
Characteristic curves
Through the selection of the Start value and End value, the proportional
output can be operated with a positive or negative characteristic curve.
Positive characteristic curve:
Select the
End value to be greater than the Start value.
The turn-on duration or frequency increases with an increasing
measured value (see page 6-12).
Negative characteristic curve:
Select the
End value to be smaller than the Start value.
The turn-on duration or frequency decreases with an increasing
measured value (see page 6-13).
The maximum values for the pulse width or switching frequency are
assigned to the End value value and the minimum values for turn-on
duration or frequency are assigned to the Start value value.
Fig. 6-6
Positive characteristic curve
Fig. 6-7
Negative characteristic curve
Measured value
Switching frequency f or
Pulse
v
width
End value
Start value
Max.
Min.
Measured value
Start value
End value
Max.
Min.
Switching frequency f or
Pulse
v
width
Summary of Contents for DIQ/S 182-XT-4
Page 6: ...Contents System 182 XT 4 0 4 ba75753e06 12 2014 ...
Page 14: ...Safety System 182 XT 4 2 4 ba75753e06 12 2014 ...
Page 68: ...Installation System 182 XT 4 3 54 ba75753e06 12 2014 ...
Page 86: ...Operation System 182 XT 4 4 18 ba75753e06 12 2014 ...
Page 92: ...The SETTINGS menu System 182 XT 4 5 6 ba75753e06 12 2014 ...
Page 128: ...Maintenance and cleaning System 182 XT 4 7 2 ba75753e06 12 2014 ...
Page 144: ...Technical data System 182 XT 4 9 12 ba75753e06 12 2014 ...
Page 146: ...Accessories and options System 182 XT 4 10 2 ba75753e06 12 2014 ...
Page 150: ...Appendix store separately if required System 182 XT 4 12 2 ba75753e06 12 2014 ...
Page 151: ......