Installation
System 182-XT-4
3 - 38
ba75753e06
12/2014
Fig. 3-27 Line power connection.
WARNING
No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages. Always cut off any wires
that are not in use as closely as possible to the cable gland.
The complete assignment of the terminal strip is shown in
section 3.13.
9
Connect phases L and N to the terminal strip. Make sure that
the cable assignment agrees with the specification on the
terminal label under the terminal strip.
10
Tighten the coupling ring (pos. 3 in Fig. 3-26).
11
Close the enclosure of the DIQ/S 182 XT-4.
N
L
Terminal
designation:
X17
X16
100...
240V AC
NETZ/MAINS
L1
N