Outputs
System 182-XT-4
6 - 26
ba75753e06
12/2014
6.5.2
Recorder
Function
The measured values of the linked sensor at the current output are set
up as current intensity in the Recorder application. The output of
measured values is laid down in the Recorder type, Start value and End
value settings.
Settings
Setting
Selection/Values
Explanation
Recorder type
0 to 20 mA or
4 to 20 mA
Start value
Within the
measuring range
(sensor-
dependent)
Minimum spacing:
≤
5 % of the measuring
range
(sensor-dependent)
End value
Measured variable
Main variable
Adjoining variable
Main variable designates
the actual measured
parameter of the sensor
(e.g. pH, oxygen, etc.).
Adjoining variable
designates an additional
measured parameter
(e.g. temperature).
Attenuation
0 ... 40 mA/s
Speed of change of the
output current (mA/s) in
the case of erratic
changes of the input
signal.
I -> OFL/UFL
Error
Current values outside
the range between Start
value and End value are
considerd as errors. The
current output reacts as
specified under Behavior
at error (see below).
Limitation
The current at the output
is limited to the Start
value or End value.
Summary of Contents for DIQ/S 182-XT-4
Page 6: ...Contents System 182 XT 4 0 4 ba75753e06 12 2014 ...
Page 14: ...Safety System 182 XT 4 2 4 ba75753e06 12 2014 ...
Page 68: ...Installation System 182 XT 4 3 54 ba75753e06 12 2014 ...
Page 86: ...Operation System 182 XT 4 4 18 ba75753e06 12 2014 ...
Page 92: ...The SETTINGS menu System 182 XT 4 5 6 ba75753e06 12 2014 ...
Page 128: ...Maintenance and cleaning System 182 XT 4 7 2 ba75753e06 12 2014 ...
Page 144: ...Technical data System 182 XT 4 9 12 ba75753e06 12 2014 ...
Page 146: ...Accessories and options System 182 XT 4 10 2 ba75753e06 12 2014 ...
Page 150: ...Appendix store separately if required System 182 XT 4 12 2 ba75753e06 12 2014 ...
Page 151: ......