2SHQLQJ'(9,&(1(70$1$*(5
Follow this procedure:
1.
Start up Windows.
2.
At the
C:
prompt, type "WIN".
3.
Double-click the
DEVICENET MANAGER
icon.
The computer interface pod has a DeviceNet node address of 62.
Nodes initially default to 00. Ensure that only one DeviceNet card
to be configured is on line with the computer.
>?D5*
A terminating resistor of approximately 121
Ω
, 1/4 W. is recommended
at the DeviceNet card.
*RLQJ2Q/LQH
To go on-line, follow these steps:
1.
Select
UTILITIES
.
2.
Select
SETUP ONLINE CONNECTION
.
3.
Establish communications with the DeviceNet card. It is
probably using 125 kB/sec. If this rate does not work, try the
other menu choices listed below. Leave the node setting at 62
for the computer interface.
>?D5*
If the computer interface pod is faulted (the Red LED is ON), turn the
interface pod switch OFF, then ON, to clear the fault.
4.
Select
YES
to go on-line. Note the
ONLINE
status at the
bottom of the main screen.
Node 62 @ 125 kB
Select GO ON LINE
then YES, GO ON LINE
Node 62 @ 500 kB
Select GO ON LINE
then YES, GO ON LINE
Node 62 @ 250 kB
Select GO ON LINE
then YES, GO ON LINE