93
ENGLISH
To carry out adjustment functions with the handset, a function key must always
be activated first before the adjustment can be carried out using the action keys
up/down (first-fault safety).
Seating position
Backkrest
Height adjustment
6.1 Basic functions of the handset
Adjustment of the backrest (individual)
Synchronous adjustment of the backrest and legrest
Height adjustment of the entire bed. This function has no influence on the position of the
patient surface.
Attention!
Since two independent drives are used for height adjustment in the sentida nursing
home bed series, the patient surface must be moved to the end position (highest or
lowest position) at least once a day, depending on the adjustment cycle. In this way,
the potential height differences of the patietn surface are compensated.
/
/
/
Please observe the warning notes on height adjustment in chapter 8!