We recommend to use mattresses manufactured with a minimum volume weight of 40 kg/m³ (density
40 kg/m³). All original wissner-bosserhoff standard mattresses fulfill this requirement. Please refer to
the following overview for the suitable mattress dimensions for the corresponding patient surfaces.
Model
Mattress dimensions
Beds with standard patient surface:
sentida 1/3/4/5/6/5-le/6-le/6-d
Aero/Easyclean patient surface 12-18 cm x 87 cm x 200 cm
Comfort patient surface 12-15 cm x 87 cm x 200 cm
Beds with a wider patient surface:
sentida 5-xl/6-xl
Aero/Easyclean patient surface 12-18 cm x 100 cm x 200 cm
Comfort patient surface 12-15 cm x 100 cm x 200 cm
Beds with wider and longer patient
surface: sentida 5-xxl
Aero-/Easyclean-Liegefläche 14-18 cm x 114 cm x 210 cm
Komfort-Liegefläche 14-15 cm x 114 cm x 210 cm
Beds with short patient surface:
sentida 5-m/6-m
Aero/Easyclean patient surface 12-18 cm x 87 cm x 190 cm
Comfort patient surface 12-15 cm x 87 cm x 190 cm
Beds with narrow patient surface:
sentida 6-s
Aero/Easyclean patient surface 12-18 cm x 80 cm x 200 cm
Comfort patient surface 12-15 cm x 80 cm x 200 cm
Beds with narrow and short patient
surface: sentida 6-xs
Aero/Easyclean patient surface 12-18 cm x 80 cm x 190 cm
Comfort patient surface 12-15 cm x 80 cm x 190 cm
Caution!
Do not use water mattresses. If the safe working load is exceeded, the drives and
other components may fail and cause accidents. Never exceed the safe working
load�
Caution!
Only replace wissner-bosserhoff mattresses with equivalent mattresses of the
same dimensions. Make sure that only mattresses with incisions on the underside
that allow for movement are used, as a mattress without these could affect the
adjustment functions or overload the drives. Do not use spring mattresses.
14.1 Mattresses