149
ENGLISH
Height adjustment motors
Electrical connection
24 V DC
Protection type
IP44
Protection class
II
Adjustment speed
max. 28 mm/s (max. 9.5 mm/s for sentida 5-le/6-le)
Compression force
max. 6,000 N (max. 8,000 N for sentida 5-le/6-le)
Patient surface motors
Electrical connection
24 V DC
Protection type
IP44
Protection class
II
Adjustment speed
6 mm/s (at the head end) 9 mm/s (at the foot end)
Compression force
2 x 6,000 N - max. lifting force (without emergency lowering)
2 x 4,500 N - max. lifting force (with emergency lowering)
2 x 3,000 N - max. tensile load
Bed weight
Approx. 115 to 130 kg depending on the model
Protection type
IPX4
Protection class
II (Protection class I for sentida 6 hospitel)
Lifter underbed clearance
14 cm (sentida 1/3/4 depending on patient surface)
Operating time
INT 2 min./18 min. The equipment is not designed for contin-
uous operation. After 2 minutes of operation at full load, a
break of 18 minutes is recommended.
Noise level, maximum
approx. 57 dB/A
Application environment
3 and 5
Humidity
30% – 75%
Atmospheric pressure
800–1060 hPa
Environment temperature
+10°C – +40°C
Manufacturer
wissner-bosserhoff GmbH
Hauptstraße 4 – 6
58739 Wickede (Ruhr)
GERMANY
Phone +49 2377 784-0