Figure 4 – 13
System Ver: It is the firmware version of the recorder. V2.20 means the present firmware
version of recorder. Plus means with the option of Math, Counter, Totalizer & CFR-21.
Boot ROM ver: It is version of Boot ROM.
Memory (Free / Total): Indicates the percentage of free memory to total memory reserved on the
recorder. 8 MB memory on board is reserved for storing measured data. A small icon on the top
right indicates the percentage of free memory e.g.: mem 96 % .
CF Card (Free / Total): Indicates the percentage of free memory to total memory of the Compact
Flash card. e.g. CF 84%.
Ethernet IP Address: It is the present IP address of the recorder.
Slot 1..6: Indicates the status of all Slots and the cards been inserted in. The cards include
Analog Input AI, Digital Input DI and Digital Output DO ( 6 relay card ).
Update: The Update key is located at left lower side. It is the key to upgrade the new firmware.
After the new firmware downloaded to the CF card, insert the CF card in. Then, press this key to
upgrade it. It may take a few seconds to finish the process.