28
Functional Verification
Welch Allyn Propaq Encore Vital Signs Monitor
-600
776.62
2400
696.36
6800
590.90
14667
434.68
-500
773.83
2500
693.80
7000
586.44
15000
428.89
-400
771.05
2600
691.26
7200
582.00
15333
423.17
-300
768.28
2700
688.72
7400
577.58
15667
417.49
-200
765.51
2800
686.18
7600
573.20
16000
411.89
-100
762.75
2900
683.66
7800
568.84
16333
406.35
0
760.00
3000
681.14
8000
564.51
16667
400.86
100
757.26
3100
678.63
8200
560.21
17000
395.44
200
754.52
3200
676.13
8400
555.93
17333
390.08
300
751.80
3300
673.64
8600
551.68
17667
384.76
400
749.08
3400
671.15
8800
547.45
18000
379.52
500
746.37
3500
668.67
9000
543.26
18333
374.34
600
743.67
3600
666.20
9200
539.08
18667
369.20
700
740.97
3700
663.73
9400
534.94
19000
364.13
800
738.28
3800
661.27
9600
530.82
19333
359.12
900
735.61
3900
658.82
9800
526.72
19667
354.15
1000
732.93
4000
656.38
10000
522.65
20000
349.24
Table 8. Local Uncorrected Barometric Pressure (LUP) as a Function of Altitude (Meters)
Altitude
(Meters)
Pressure
(mmHg)
Altitude
(Meters)
Pressure
(mmHg)
Altitude
(Meters)
Pressure
(mmHg)
Altitude
(Meters)
Pressure
(mmHg)
-700
825.23
1000
674.11
2700
546.23
4400
438.74
-650
820.43
1050
670.03
2750
542.79
4450
435.86
-600
815.65
1100
665.98
2800
539.37
4500
432.99
-550
810.89
1150
661.94
2850
535.96
4550
430.14
-500
806.15
1200
657.92
2900
532.58
4600
427.31
-450
801.43
1250
653.92
2950
529.21
4650
424.49
-400
796.74
1300
649.94
3000
525.85
4700
421.69
-350
792.07
1350
645.98
3050
522.52
4750
418.90
-300
787.43
1400
642.04
3100
519.20
4800
416.12
-250
782.80
1450
638.12
3150
515.90
4850
413.36
-200
778.20
1500
634.22
3200
512.62
4900
410.62
-150
773.61
1550
630.34
3250
509.35
4950
407.89
-100
769.05
1600
626.48
3300
506.10
5000
405.18
-50
764.52
1650
622.63
3350
502.87
5050
402.48
0
760.00
1700
618.81
3400
499.65
5100
399.79
50
755.51
1750
615.00
3450
496.45
5150
397.12
100
751.03
1800
611.22
3500
493.27
5200
394.46
150
746.58
1850
607.45
3550
490.10
5250
391.82
200
742.15
1900
603.70
3600
486.95
5300
389.19
250
737.74
1950
599.97
3650
483.82
5350
386.58
300
733.35
2000
596.26
3700
480.70
5400
383.98
350
728.99
2050
592.57
3750
477.60
5450
381.39
400
724.64
2100
588.89
3800
474.51
5500
378.82
450
720.32
2150
585.24
3850
471.44
5550
376.27
500
716.01
2200
581.60
3900
468.39
5600
373.72
550
711.73
2250
577.98
3950
465.36
5650
371.20
600
707.47
2300
574.38
4000
462.33
5700
368.68
650
703.23
2350
570.80
4050
459.33
5750
366.18
700
699.01
2400
567.24
4100
456.34
5800
363.69
750
694.81
2450
563.69
4150
453.37
5850
361.22
800
690.63
2500
560.16
4200
450.41
5900
358.76
850
686.47
2550
556.65
4250
447.47
5950
356.31
900
682.33
2600
553.16
4300
444.54
6000
353.88
950
678.21
2650
549.68
4350
441.63
6050
351.46
Table 7. Local Uncorrected Barometric Pressure (LUP) as a Function of Altitude (Feet)
Altitude
(Feet)
Pressure
(mmHg)
Altitude
(Feet)
Pressure
(mmHg)
Altitude
(Feet)
Pressure
(mmHg)
Altitude
(Feet)
Pressure
(mmHg)
Summary of Contents for Propaq Encore
Page 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Page 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Page 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Page 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Page 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Page 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...