B
125
Dynatech/Nevada Patient Simulator
Modification
The factory configuration of the Dynatech/Nevada model 214A, 215A, and 217A Patient
Simulators is incompatible with the Propaq Encore monitors with IBP channels when both
the ECG and invasive pressure simulator channels are simultaneously connected to the
monitor. Erroneous readings result. The patient simulator must be modified as described
here before you can use both simulator channels with the Propaq Encore monitors with
IBP. If only one simulator channel is used, the modification is unnecessary.
Modify the patient simulator's invasive pressure cable as shown in the figure below.
Remove jumpers W1 and W2 from the patient simulator's Top PCB.
Incompatibilities with other simulators may also cause erroneous readings and/or missing
ECG. Check Propaq Encore readings with the simulator against a known source before
using the simulator for calibration.
Note
Without this modification, the monitor will select a single waveform or parameter
for display.
Function
Color
Pin No.
Encore 2X4/2X6
Output (+)
Black
4
C
Output (-)
Red
1
A & E
Exciter (+)
White
3
B
Exciter (-)
Green
5
D
ECG Ref
Blue
2
No Connection
3
5
2
4
1
810 0696
W1
W2
Summary of Contents for Propaq Encore
Page 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Page 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Page 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Page 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Page 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Page 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...