Service Manual
Calibration
39
8.
Disconnect the DMM leads.
Calibrating Realtime ECG Out
1.
Connect the 5-lead ECG cable to the monitor and the simulator.
2.
Set the simulator to 80 bpm, normal sinus rhythm.
3.
Set the monitor to Lead II.
4.
Connect the dc-coupled oscilloscope probe to the
Digital board
at P9 pin 6, and the
reference probe to TP4 on the
Digital board
.
5.
On the Analog board, carefully short together U55 pins 1 and 2.
6.
Adjust RP6 on the Analog board to 0V dc ± 0.1V.
7.
Remove the short and check for an ECG signal displayed on the oscilloscope of 1.0V±
0.1V baseline to peak of the R-wave.
8.
Disconnect the ECG cable from the monitor.
+VB
TP43 or TP1
TP13
+5.40 to +6.10
+5.6
TP43 or TP1
TP33
+5.55 to +5.70
IBP shield
TP43 or TP1
P6, Pin 10
+2.47 to +2.53
C.L.P. detector
a
TP43 or TP1
U5, Pin 1
0.00 to +0.25
NIBP Pressure Offset
TP43 or TP1
TP31
+2.40 to +2.60
OSC baseline
TP43 or TP1
TP47
-0.95 to -1.25
+VA
TP43 or TP1
TP7
(+VR + 0.1V) to (+VR + 0.2V)
-VA
TP43 or TP1
TP15
(+VR + 0.2V) to -(VR + 0.35V)
+2.5
TP43 or TP1
TP17
+2.48 to+2.52V
a.
Make sure the ECG input connector (P5) is disconnected from the analog board for this
measurement. Reconnect when done.
Table 11. Main Power Supply Voltage Checks
Supply Voltages
Ref. Lead
Pos. Lead
Voltage Limits
P9
PIN 6
TP4
GND
Summary of Contents for Propaq Encore
Page 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Page 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Page 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Page 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Page 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Page 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...