Service Manual
Repair Procedures
91
Replacing Expansion Module Side Panels
The side panels easily slide out after the chassis and rear panel are separated. If the
patient connector side panel is being replaced, this side panel and the two circuit boards
must be removed together before disconnecting the side panel from the circuit board.
See
“Replacing the SpO2 Circuit Boards”
on page 86 or
If the speaker side panel is being replaced, separate the chassis and rear panel as
described earlier in this section, disconnect the speaker from the SCP or MSP board, and
slide out the speaker side panel.