Service Manual
Repair Procedures
75
Replacing the Side Panels
The Propaq Encore side panels are complete assemblies.
The side panels are sealed to the rear chassis with an adhesive gasket. A new gasket is
provided with the new side panel kit.
Removing the Right Side Panel
1.
Remove the battery pack (
2.
Open the monitor (
3.
Remove the Recharger board (
“Replacing the Recharger Board”
4.
Cut and remove the tie wrap securing the bead connected to the side panel.
5.
Remove the two right panel clamps by inserting the end of the flat blade screwdriver
under the clamp and gently rotating the blade to loosen the clamp. Then slide the
clamp out.
6.
Press the side panel out of the rear chassis. There is an adhesive gasket around the
side panel which requires considerable force to separate.
Do not pry against
components
.
Removing the Left Side Panel
1.
Remove the battery pack (
2.
Open the monitor (
3.
Remove the Recharger board (
“Replacing the Recharger Board”
4.
Remove the U-shaped clamp that secures the left side panel assembly by inserting a
small screwdriver between the clamp and plastic and the end of another flat blade
screwdriver under the clamp and gently rotating the blade to loosen the clamp. Then
slide the clamp out.
0696-42
INSERT SMALL SCREWDRIVER
BETWEEN PLASTIC AND CLAMP
TO RELIEVE TENSION AGAINST
PLASTIC
Summary of Contents for Propaq Encore
Page 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Page 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Page 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Page 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Page 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Page 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Page 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...