Requirement:
A contact closure output connection (Inject Start Out) from the sample manager
is required to trigger a mass spectrometer running under MassLynx software control to start.
Figure A–9: Sample manager I/O signal connectors
Inject Start Out +
Inject Start Out -
Ground
Ground
Inject Hold In +
Inject hold in -
For electrical specifications, see the
µSample Manager - Fixed Loop Overview and Maintenance
Guide
on your documentation media, or at
, for more details.
Table A–3: Sample manager event-out/event-in connections
Signal connections Description
Inject Start Out
Indicates (with a contact closure output) that an injection has started.
Inject Hold In
Delays the next injection when the sample manager receives a contact
closure input (from another system module, for example).
A.7.3 TUV detector signal connectors
If your system includes a TUV detector, consult the
ACQUITY UPLC TUV Detector Operator's
Overview and Maintenance Guide
on your documentation media, or at
, for more
details on signal connectors.
A.7.4 PDA detector signal connectors
If your system includes a PDA detector, consult the
ACQUITY Photodiode Array Detector with
TaperSlit Overview and Maintenance Guide
on your documentation media, or at
, for
more details on signal connectors.
April 4, 2018, 715005704 Rev. A
Page 66