Figure A–5: Positive and negative lead connections
Screw
Connector
Signal cable
3.
Fit the grounding cable's fork terminal on the rear-panel grounding stud and secure the
terminal using the locking nut.
Note:
Use the 9/32-inch nut driver to tighten the locking nut until the terminal does not
move.
Figure A–6: Grounding cable fork terminal on grounding stud
Fork terminal
Locking nut
Grounding stud
April 4, 2018, 715005704 Rev. A