3.5.1 Binary solvent manager control panel
The binary solvent manager control panel displays flow status, system pressure, total flow rate,
and solvent composition parameters.
Rule:
You can edit these parameters when the system is idle by clicking the underlined value.
You cannot edit binary solvent manager parameters while the system is running samples.
Figure 3–1: Binary solvent manager control panel
The following table describes the items in the binary solvent manager control panel:
Flow LED
– Displays the actual flow LED on the front panel of the binary solvent manager
unless communications with the binary solvent manager are lost.
Solvent composition
– Displays the percentage of solvent to be drawn from the solvent
lines (A and B). Composition values range from 0.0 to 100.0%.
Stop flow
– Immediately stops all flow from the binary solvent manager.
Pressure Delta –
The difference between the maximum and minimum pressures observed
during the previous minute of operation.
Flow rate
– Displays the total flow rate of the binary solvent manager, from 0.000 to 2.000
mL/min under normal operation and 0.000 to 8.000 mL/min when priming.
System pressure
– Displays system pressure, in kPa, bar, or psi. You can customize
pressure units via the ACQUITY UPLC Console.
Status
– Displays the status of the current operation.
You can access these additional functions by right-clicking anywhere in the binary solvent
manager control panel:
Table 3–4: Additional functions in the binary solvent manager control panel
Control panel function Description
Start up system
Brings the system to operational conditions after an extended idle
period or when switching to different solvents.