2-28
Setting up the Detector
Inject start connections between the Alliance separations module and the
detector
Generating stop flow
The detector has a dedicated switch output, called “stop flow” that becomes
active when a heater, gas flow, or lamp failure occurs.
To use the stop flow function, make the connections shown in the following
table and figure.
Warning:
Fire and explosion hazard. Failure to connect the stop flow
output could cause the detector to flood.
1
2
3
4
5
6
7
8
9
10
+
−
+
−
+
−
+
−
Inject Start In
Inject Start In
Ground
Lamp On
Lamp On
Chart Mark In
Chart Mark In
Ground
Auto Zero In
Auto Zero In
1
2
3
4
5
6
7
8
9
10
11
12
+
−
+
−
+
−
+
−
Inject Start
Inject Start
Ground
Stop Flow
Stop Flow
Hold Inject 1
Hold Inject 1
Hold Inject 2
Hold Inject 2
Ground
Chart Out
Chart Out
Alliance connector B
2424 ELS detector connector II
Red
Black