5-30
Diagnostic Functions and Troubleshooting
Use the following table to troubleshoot peak resolution problems that may be
affecting your results.
Peak resolution troubleshooting
Symptom
Possible cause
Corrective action
Straight baseline, no
peaks
No pump flow
Set pump flow rate.
Lamp not operating
Check method. Lamp might be
turned off.
Call Waters Technical Service.
Detector not zeroed
Auto zero detector baseline.
Check cables between unit and
recorder.
Improper connection
between the detector
and the recorder
Auto zero detector baseline.
Check cables between unit and
recorder.
No sample injected
Check injector.
Leak in solvent path Check fittings and drip tray.
Bad column
Clean or flush or replace
column.
Call Waters Technical Service.
Detector gas flow set
to off
Turn gas flow back on.
No gain set
Set gain (see
Flat-topped peaks
Detector not zeroed
Auto zero detector baseline.
Incorrect recorder
input voltage
Adjust recorder input voltage,
or connect cable to correct port
on recorder.
Sample
concentration or
injection volume
exceeds voltage
output of detector
Decrease sample concentration
or injection volume.