5-32
Diagnostic Functions and Troubleshooting
Increased noise
(continued)
Drift tube
temperature too low.
Increase drift tube
temperature (see
).
Drift tube
temperature too
high.
Decrease drift tube
temperature (see
).
Nebulizer
temperature too
high.
Decrease nebulizer power level
(see
).
Incorrect results troubleshooting (Continued)
Symptom
Possible cause
Corrective action