2-14
Setting up the Detector
Connecting the siphon drain tubing
Siphon waste is routed down the front or to the rear of the detector via one of
two different siphon drain tubes supplied in the startup kit.
Tip:
At initial setup, Waters recommends that you fill the siphon with water
or mobile phase. Omitting this step delays satisfactory detector performance.
Routing the siphon drain tubing down the front of the detector
Required materials
•
Front siphon drain tubing (in startup kit)
•
Hose clamp, to secure front siphon drain tubing to siphon (in startup kit)
To route the siphon drain tubing down the front of the detector
1. Slide the 0.68 × 0.74 cm (.269 × .291 inches) hose clamp onto the end of
the siphon drain tubing.
Hose clamp on end of siphon drain tubing
Hose clamp
End of siphon
drain tubing