Connecting the siphon drain tubing
2-15
2. Attach the front siphon drain tubing to the stainless steel siphon drain
on the front of the detector and pinch the clamp so that it is snug.
Front siphon drain tubing installation
3. Ensure that the siphon drain tubing is routed to a collection bottle in
such a way that condensed solvent can flow freely from the condenser
and the end of the tubing is not immersed in the collected liquid.
Routing the siphon drain tubing to the rear of the detector
Tip:
The rear siphon drain tubing is intended for use with the optional Waters
Alliance bottle tray kit.
Required material
•
Rear siphon drain tubing (in startup kit)
•
Convoluted tubing (in startup kit)
•
Plastic hose clamp (in startup kit)
•
Siphon drain tubing bracket (in startup kit)
•
Phillips
®
screwdriver
Warning:
To avoid flooding the detector, ensure the siphon drain
tubing is not kinked or submerged in the collection bottle.
Hose clamp
Siphon drain tubing