Loading the chamber
Loading the chamber
Loading the chamber
Loading the chamber
Tubes
Tubes
Tubes
Tubes
Rinse, drain and dry tubes after washing.
Place tubes on a tray allowing the ends to remain open. Do not bend tubes.
Wrapped/bagged items
Wrapped/bagged items
Wrapped/bagged items
Wrapped/bagged items
Place the bags on trays allowing adequate space in-between bags.
Ensure that packs do not touch the sterilizer chamber walls.
Place sterilization bags with the paper side facing up.
Paper side up
Space between bags
Before initiating a sterilization cycle, always check that the steam diffuser plate is properly
positioned.
Never place the load or the trays directly into the chamber without the chamber rack as
this could affect the steam and temperature distribution. The load must always be
supported by the chamber rack.
Before touching, ensure the sterilization chamber is cold: risk of burns!
An improper positioning of the steam diffuser plate could result in bad steam
quality and could impair the sterilization process, with risk of non sterile load and
cross infection.
Sterility at the end of the cycle is not guaranteed if the steam diffuser plate was
not correctly placed.