Required minimum clearances
Required minimum clearances
Required minimum clearances
Required minimum clearances
Back side:
50mm
Right and left sides:
10 mm
Upper side:
As required for filling the water tank, 50 mm minimum
Placement
Placement
Placement
Placement
Place the sterilizer on a flat and level surface, far from sources of heat and from flammable materials.
Do not place the sterilizer so that it is difficult to open the service door and operate on the controls in it.
Do not place the sterilizer so that it is difficult to disconnect the power supply plug.
Place the sterilizer in a well ventilated room.
If installed in a cabinet, this shall be provided with an opening of at least 200x150 mm on the rear side.
The sterilizer must not be operated in presence of explosive atmospheres.
Electrical connection
Electrical connection
Electrical connection
Electrical connection
The electrical power supply to the sterilizer must fulfil all applicable standards in the country of use,
and must comply with the data label on the back of the sterilizer.
Connect the cord set to the socket provided in the back of the sterilizer.
Connect the mains cable to a wall outlet with the following characteristics:
•
Single - phase 200 - 240 V, 50/60 Hz, 8,75 A, on a dedicated circuit;
•
Overvoltage category = II;
•
10 A differential circuit breaker with a sensitivity of 30 mA. The circuit breaker must be a certified
type according to applicable norms;
•
Maximum power consumption of the sterilizer is 1750 W;
•
A grounded connection is essential.