38
Maintenance program
Maintenance program
Maintenance program
Maintenance program
MAINTENANCE PROGRAM TABLE
MAINTENANCE PROGRAM TABLE
MAINTENANCE PROGRAM TABLE
MAINTENANCE PROGRAM TABLE
Frequency (*)
# of cycles (*)
Operation
Consumable
Performed by
Monthly
50
Clean the door seal and the chamber face side
See
APPENDIX 8
User
Clean the chamber, trays and the rack
Clean the chamber filter
Clean the external sterilizer surfaces
Clean the steam diffuser plate
3 months
400
Replace the bacteriological filter
Replace the dust filter
6 months
800
Clean both water tanks
Yearly
800
Replace the door seal
5 years
4000
General check and service
Service technician
-
20000
General check and service
(*) whichever occurs first