96
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
Technical data
Technical data
WATER SUPPLY SYSTEM
Temperature
max. 35 °C (95 °F)
Pressure
min. 2 bar – max. 8.6 bar (min 29 psi - max. 124.7 psi)
Flow
min. 0.25 – max. 0.5 l/min (min. 0.066 - max. 0.132 gal/min)
POWER SUPPLY SYSTEM
Nominal voltage and Max.
current
200–240 V ac (±10%), 50 Hz, 10 A, single-phase
200–240 V ac (±10%), 60 Hz, 10 A, single-phase
Overvoltage category
II
Protection required
Suitable circuit breaker and a Ground Fault Circuit Interrupter (GFCI).
All protection devices must be certified according to applicable
standard. A grounded connection is essential.
Communication with other
devices
1 USB port in the front
1 USB port at the rear (optional)
Features
Fully micro-processor controlled, process evaluation system
according to EN13060.
Programmable standby mode.
Max. heat output
3000 kJ/h
INSTALLATION REQUIREMENTS
Working temperature
From +5 °C to +40 °C (from +41 °F to +104 °F)
Working relative humidity
Max. RH 80% up to 31 °C (88 °F), linearly decreasing to 50% at 40
°C (104 °F)
Storage temperature / rel.
humidity
From -20 °C to +60 °C (from -4 °F to +140 °F )/0–90 % (with empty
tanks)
Max altitude
3000 m asl
Min. atmospheric pressure
0.6 bar (8.7 psi)
Overall dimensions
W: 465 mm/H: 452 mm/D: 646 mm (W: 18.3"/H: 17.8"/D: 25.4")
Min. space required
W: 485 mm/H:502 mm/D:439 mm (W: 19,1"/H: 19,8"/D: 17,3")
Size of the door movement
W: 532 mm/H: 443 mm/D: 357 mm (W: 21"/H: 17,5"/D: 14")
Weight empty
LINA 17: 40 kg (88.2 lbs)
LINA 22: 42 kg (92.6 lbs)
Max. weight (fully loaded)
LINA 17: 54 kg (119 lbs)
LINA 22: 56.5 kg (124.6 lbs)
Weight per support area
LINA 17: 32 kN/m
2
LINA 22: 33.5 kN/m
2
Environment pollution
Degree 2
Usage environment
Indoor
STERILIZER CHAMBER
Pressure safety valve
2.6 bar (37.7 psi)
Safety thermostats
180 °C (356 °F)
Total volume
LINA 17: 17 l/Ø: 250 mm/D: 362 mm (4.5 gal, Ø: 9.8’’/D: 14’’)
LINA 22: 22 l/Ø: 250 mm/D: 440 mm (5.8 gal, Ø: 9.8’’/D: 17’’)
Usable space *
LINA 17: W: 195 mm/H: 195 mm/D: 312 mm (W: 7.7’’/H: 7.7’’/D:
11.7’’)
LINA 22: W: 195 mm/H: 195 mm/D: 400 mm (/W: 7.7’’/H: 7.7’’/D:
15.4’’)
Bacteriological filter
0.3 µm
Note*: Usable space with standard rack and trays. With optional racks and