Maintenance
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
65
PREPARE THE STERILIZER FOR CLEANING THE WATER TANKS
Proceed as follows:
1
Switch OFF the sterilizer and disconnect the mains cable.
2
Completely drain both water tanks (see "Draining the used and clean water tank" on page 73).
3
To drain the detergent solution during the subsequent cleaning, leave the drain tube attached to the drain port of the tank you want to clean.
ACCESS TO THE WATER TANKS
Proceed as follows:
1
Lift the water tank cover.
Summary of Contents for lina
Page 1: ...RIK 117 RIK 122 Instructions for Use RIK 1 Med ENG Rev03 ...
Page 2: ......
Page 113: ......