Maintenance
70
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
800-cycle maintenance
REPLACING THE DOOR GASKET
Notice: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page 56.
Proceed as follows:
1
Fully open the chamber door.
2
Remove the used door gasket by hand.
3
Carefully clean the seal seat and the chamber face side with a damp, lint-free
cloth.
4
Moisten the new seal with water to facilitate its placement.