56
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
Maintenance
Frequency
1
Cycles
1
Operation
Yearly
2
400
2
Replace the bacteriological filter. See "400-cycle
maintenance" on page 62.
Replace the dust filter. See "400-cycle
maintenance" on page 62
Yearly
2
800
2
Replace the door gasket. See "800-cycle
maintenance" on page 70.
5 years
4000
General check and service. See "4000 cycle or
five-year maintenance" on page 72.
Note
1
: Whichever occurs first.
Note
2
: Even if the maximum cycle number is not reached, it is
recommended to replace the consumable parts every year, or if they appear
worn or damaged, or if the filters are clogged or discolored.
EXPIRED MAINTENANCE
The sterilizer monitors the wear of consumables by counting the
number of cycles executed since the last replacement.
When the number of cycles is close to the maximum, a pre-alert
about the concerned consumable is displayed. Please check that
you have the requested spare part available, buy one if not. When
the maximum number of cycles has been met, a message to
replace the consumable will be displayed.
If you can not replace the consumable immediately, the sterilizer
will operate anyway but the message will appear again some cycles
later.
When you have replaced the consumable tap
to confirm: the
executed cycle counter is reset.
REPLACE THE CONSUMABLE BEFORE THE MAINTENANCE DUE
DATE
If you replace the consumables before the request of replacement
appears, you should manually reset the counters through the
following procedure.
1
On the homepage tap
>
2
Select the consumable you want to replace: a message appears
showing the current worked hours of the part.
3
When you have replaced the consumable tap
to confirm: the
executed cycle counter is reset.