Sterilizer tests
40
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
4
Place the tube with the capsule in the center of a tray in the lowest rack position
and close the chamber door.
5
On the homepage tap Helix-B&D test.
6
To set the duration of the Plateau/Sterilization phase and other settings, tap
.
7
Tap
and enter your credentials if required: the chamber door locks.
8
Wait until the end of the test and tap OPEN: the chamber door unlocks.
9
Enter your credentials if required.
10
Open the chamber door, extract the tray using the tray holder and take the tube.
CAUTION! Risk of burns. The test pack is very hot at the end of the cycle.
Wear appropriate PPE (e.g. gloves).
11
Unscrew the tube capsule and remove the indicator strip.
12
Check the change in colour. See "Interpret the test result" below
INTERPRET THE TEST RESULT
Indicator
What happened
Test
passed
What to do next
The indicator strip has turned dark. The air was completely removed from
the capsule.
Yes
-
Part of the chemical indicator strip has not turned dark. The air removal
from the capsule was not complete.
No
Repeat the test. If it fails repeatedly,
call technical service.