Maintenance
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
63
REPLACING THE DUST FILTER
Notice: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page 56.
Proceed as follows:
1
Open the chamber door.
2
Pull out the dust filter handle from underneath the sterilizer.
3
Lift the used filter from the handle and remove it.