18
LINA | Instructions for Use | RIK-1 Med ENG Rev03 | 11/05/2020 | © 2020 W&H Sterilization Srl
Getting started
Installing the sterilizer
LOCATION REQUIREMENTS
Notice:
Do not place the sterilizer so that it is difficult to operate the
controls behind the service door. Do not place the sterilizer so that
it is difficult to disconnect the power supply plug.
Leave the condenser grid (rear side of the sterilizer) free from
anything that might obstruct the air passage.
Surface materials should be water resistant. If sterilization cycles
will be continuous, pay attention to the surrounding materials:
steam can damage them.
The sterilizer must operate in absence of explosive atmospheres.
The sterilizer must operate in a well ventilated room (indoor), far
from sources of heat and from flammable materials.
Place the sterilizer on a flat and level surface.
Clearance requirements to ensure proper air circulation:
ELECTRICAL CONNECTIONS
All the cables and tubes connected on the rear side of the sterilizer
must be placed far from the condenser grid (e.g. using the available
guides).
Notice:
Connect the sterilizer to a dedicated line. Do not use cable
extensions nor multiple sockets/adapters.
Ensure that external and internal surfaces are free from moisture or
condensation before connecting to power.