ORP voltage
MU 6100 H
38
ba77077e02
09/2014
7
ORP voltage
7.1
Measuring
7.1.1
Measuring the ORP
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect
.
The USB interface is not galvanically isolated.
Stability control
(AutoRead )
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 10.2.3 A
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
1.
Connect the ORP electrode to the meter.
2.
If necessary, select the U (mV) display with
<MODE>
.
3.
When measuring without temperature sensor:
– Temper the test sample, or measure the current temperature.
– Enter the temperature value in the menu.
4.
Rinse the ORP electrode and immerse it in the test sample.
The measured value is checked for stability (automatic stability control).
The display of the measured parameter flashes.
5.
Wait for a stable measured value.
The display of the measured parameter no longer flashes.
U
157.0
mV
24.8 °C
15.02.2014
14:15
Menu
Measured
parameter
Time interval
Stability in the time interval
ORP
15 seconds
Δ
: better than 0.3 mV
Temperature
15 seconds
Δ
: better than 0.5 °C