Settings
MU 6100 H
54
ba77077e02
09/2014
* Brand names or trade names are trademarks of their respective owners protected
by law.
10.1.3 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
Setting the calibra-
tion interval
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
5
Merck 1*
4.000
7.000
9.000
20°C
6
Merck 2 *
1.000
6.000
8.000
13.000
20°C
7
Merck 3 *
4.660
6.880
9.220
20°C
8
Merck 4 *
2.000
4.000
7.000
10.000
20°C
9
Merck 5 *
4.010
7.000
10.000
25 °C
No.
Buffer set *
pH values
at
The buffers are selected in the menu, pH /
<OK>
/
Calibration
/
Buffer
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
1.
Open the menu for measurement settings with
<OK>
.
2.
In the
Calibration
/
Calibration interval
menu, set the calibration interval
with
<
><
>
.
3.
Confirm the setting with
<OK>
.
4.
Quit the menu with
<MODE>
.