MU 6100 H
Technical data
ba77077e02
09/2014
11
3.1.2 Oxi
Measuring ranges,
resolution
(depending on the
sensor)
Variable
Measuring range
Resolution
Concentration [mg/l]
0 ... 20.00 (0 ... 20.0)
20.0 ... 90.0
(20 ... 90)
0.01 (0.1)
0.1 (1)
Saturation [%]
0 ... 200.0 (0 ... 600)
0 ... 600
0.1 (1)
1
D.O. partial pressure
[mbar]
0 ... 200.0 (0 ... 1250)
0 ... 1250
0.1 (1)
1
T [°C]
0 ... 50,0
0.1
Accuracy
(± 1 digit)
Variable
Accuracy
Concentration [mg/l]
± 0.5 % of measured value
at ambient temperature of +5 °C ... +30 °C
Saturation [%]
± 0.5% of measured value
when measuring in the range of ± 10 K
around the calibration temperature
D.O. partial pressure
[mbar]
± 0.5 % of measured value
at ambient temperature of +5 °C ... +30 °C
T [°C] / temperature sensor
NTC 30
± 0.1
PT 1000
± 0.1
Correction functions
Temperature
compensation
Accuracy better than 2 % at 0 ... +40 °C
Salinity correction
0 ... 70.0 SAL
Air pressure correc-
tion
Automatic through integrated
pressure sensor in the range of 500 ... 1100
mbar
The accuracy values specified here apply exclusively to the meter.
The accuracy of the D.O. sensors has also to be taken into
account.