Dissolved oxygen
MU 6100 H
40
ba77077e02
09/2014
8
Dissolved oxygen
8.1
Measuring
8.1.1
Measuring D.O.
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect
.
The USB interface is not galvanically isolated.
Salinity correction
When measuring the concentration of solutions with a salt content of more than
1 g/l, a salinity correction is required. For this, you have to measure and input
the salinity of the measured medium first. When the salinity correction is
switched on, the
[SAL]
indicator is displayed in the measuring window.
Stability control
(AutoRead )
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 10.2.3 A
1.
Connect the D.O. sensor to the meter.
The D.O. measuring screen is displayed.
2.
If necessary, select the measured parameter with
<MODE>
.
3.
Immerse the D.O. sensor in the test sample.
4.
If necessary, select the measured parameter with
<MODE>
.
D.O. concentration [mg/l]
D.O. saturation [%]
D.O. partial pressure [mbar].
O
2
7.9
mg/l
25.1
°C
[OXY 11]
USB output
15.02.2014
08:00
You can switch the salinity correction on or off and enter the salinity
in the menu for calibration and measurement settings (see section
10.1.5 S