What to do if...
MU 6100 H
80
ba77077e02
09/2014
No stable measured
value
Obviously incorrect
measured values
14.2 Dissolved oxygen
Error message
OFL, UFL
The measured value is outside the measuring range.
Cause
Remedy
Electrode:
– Junction contaminated
– Clean the junction
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Temper if necessary
Ele test sample:
– Conductivity too low
– Use a suitable electrode
– Temperature too high
– Use a suitable electrode
– Organic liquids
– Use a suitable electrode
Cause
Remedy
Electrode:
– Electrode unsuitable
– Use a suitable electrode
– Temperature difference between
buffer and test sample too great
– Adjust temperature of buffer or
sample solutions
– Measurement procedure not suit-
able
– Follow special procedure
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– The measured value is outside
the measuring range of the meter
– Use a suitable D.O. sensor