MU 6100 H
Settings
ba77077e02
09/2014
59
Measurement
/
Measuring cell
/
Type
Measuring cell
used
Cal
Measuring cells whose cell constant
is determined by calibration in the
KCL control standard solution.
Calibration ranges:
0.450 ... 0.500 cm
-1
and
0.800 ... 0.880 cm
-1
The currently valid cell constant is
displayed in the status line.
C 0.100
Measuring cell C 0.100,
nominal cell constant
0.100 cm
-1
.
The cell constant can be adjusted in
the range 0.090 ... 0.110 cm
-1
.
C 0.010
Measuring cell C 0.010,
nominal cell constant
0.010 cm
-1
.
The cell constant is permanently set.
man
Any measuring cells with freely
adjustable cell constants.
Measurement
/
Measuring cell
/
Cell const. man
0,250 ...
0,840
1,000 cm
-1
Display and setting option of the cell
constant of any measuring cells
(man).
Measurement
/
Measuring cell
/
Cell const. LR325/
01
0.090 ...
0.100
...
0.110 cm
-1
Display and setting options for the cell
constant of the measuring cell
C 0.100.
Measurement
/
Temp. comp. (TC)
/
Method
nLF
Lin
Off
Procedure for temperature compen-
sation (see section 10.1.9 T
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
Measurement
/
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
10.000 %/K
Coefficient of the linear temperature
compensation.
This setting is only available when the
linear temperature compensation is
set.
Menu item
Possible
setting
Explanation