www.vtiinstruments.com
VM2710A Command Dictionary
151
TRACe:FEED
Purpose
Sets data retrieval method
Type
Setting
Command Syntax
TRACe:FEED <type>
Command Parameters
<type> = OFF | REGister
*RST Value
OFF
Query Syntax
TRACe:FEED?
Query Parameters
N/A
Query Response
OFF | REG
Description
Data (measurements) are retrieved from the VM2710A using two different methods.
For most systems, the easiest method to use is word serial data transfers. This is the
same technique used to program the module and query the module's settings. When fast
response time is important, measurements can be retrieved using register access. The
two methods are exclusive; only one can be used at a time.
Setting TRACe:FEED to OFF is the way to use word serial data transfers; this is the
unit default. Setting TRACe:FEED to REGister is the way to use register access.
Note that in register access mode the MATH and CALCulate functions are not available
with the exception of CALCulate:[FUNCtion:]LIMit.
See Section 3 for more information on register access.
Examples
Command / Query
Response (Description)
TRAC:FEED REG
(Sets the unit for register interface.)
Related Commands
N/A