VTI Instruments Corporation
20
VM2710A Introduction
R
ESISTANCE
S
PECIFICATIONS
(C
ONTINUED
)
4-W
IRE
:
90
D
AY
A
CCURACY
±(%R
EADING
+
%R
ANGE
)
T
CAL
±5
°C
AFTER
1
H
R
W
ARM
-
UP
3
Range
100 PLC
10 PLC
1 PLC
0.1 PLC
0.03 PLC
20 Ω
0.005 + 0.008
0.005 + 0.008
0.005 + 0.008
0.005 + 0.010
0.02 + 0.02
200 Ω
0.005 + 0.008
0.007 + 0.008
0.007 + 0.008
0.007 + 0.010
0.02 + 0.02
2 kΩ
0.005 + 0.002
0.007 + 0.003
0.007 + 0.005
0.007 + 0.007
0.02 + 0.02
20 kΩ
0.005 + 0.003
0.007 + 0.003
0.007 + 0.005
0.007 + 0.007
0.02 + 0.02
200 kΩ
0.005 + 0.003
0.007 + 0.003
0.007 + 0.005
0.007 + 0.007
0.02 + 0.02
2 MΩ
0.010 + 0.003
0.010 + 0.003
0.010 + 0.005
0.050 + 0.007
0.04 + 0.03
20 MΩ
0.020 + 0.003
0.020 + 0.003
0.020 + 0.005
0.200 + 0.007
0.20 + 0.05
4-W
IRE
:
1
Y
R
A
CCURACY
±(%R
EADING
+
%R
ANGE
)
T
CAL
±5
°C
AFTER
1
H
R
W
ARM
-
UP
3
Range
100 PLC
10 PLC
1 PLC
0.1 PLC
0.03 PLC
20 Ω
0.010 + 0.010
0.010 + 0.010
0.010 + 0.010
0.010 + 0.015
0.03 + 0.03
200 Ω
0.010 + 0.010
0.010 + 0.010
0.010 + 0.010
0.010 + 0.010
0.03 + 0.03
2 kΩ
0.010 + 0.005
0.010 + 0.005
0.010 + 0.007
0.010 + 0.007
0.03 + 0.03
20 kΩ
0.010 + 0.005
0.010 + 0.005
0.010 + 0.007
0.010 + 0.007
0.03 + 0.03
200 kΩ
0.010 + 0.005
0.010 + 0.005
0.010 + 0.007
0.010 + 0.007
0.03 + 0.03
2
MΩ
0.015 + 0.005
0.015 + 0.005
0.015 + 0.010
0.060 + 0.010
0.05 + 0.04
20 MΩ
0.025 + 0.005
0.025 + 0.005
0.025 + 0.010
0.210 + 0.010
0.22 + 0.05
T
EMPERATURE
C
OEFFICIENT
;
0
°C
-
18
°C
AND
28
°C
-
55
°C;
±(%R
EADING
+
%R
ANGE
)
/
°C
Range
20 Ω
200 Ω
2 kΩ
20 kΩ
200 kΩ
2 MΩ
20 MΩ
0.0007 + 0.0002
0.0007 + 0.0002
0.0006 + 0.0001
0.0006 + 0.0002
0.0006 + 0.0002
0.0015 + 0.0004
0.0035 + 0.0004
R
EADING
R
ATE
S
PECIFICATIONS
R
ESISTANCE
R
ANGES
Read Rates (60 Hz)
Readings/s
Aperture Times
Read Rates (70 Hz)
Readings/s
Aperture Times
0.5
5.0
50.0
500.0
2000.0
100 PLC 10 PLC
1 PLC
0.1 PLC
0.03 PLC
0.4
4.0
45.0
450.0
2000.0
100 PLC
10 PLC
1 PLC
0.1 PLC
0.03 PLC
Note
1
Accuracy @ 100, 10 and 1 PLC @ T
CAL
±1 °C after 1 hr warm-up; ±(%R %Range), Sine-wave input
≥
5% F.S., Low-Frequency Filter ON
2
Accuracy @ 100, 10 and 1 PLC @ T
CAL
± 5 °C after 1 hr warm-up; ±(%R %Range), Sine-wave input
≥
5% F.S., Low-Frequency Filter ON
3
2-wire accuracy is the same as that shown for 4-wire measurements after performing a math null measurement. Without setting the null, readings will
include lead resistance, which is application specific, and add 0.2
Ω of additional error for contact resistance. Typical 24 AWG stranded wire is
0.03
Ω/ft.